Air Jacketed Co2 Incubator LX400AJI

Description

Air Jacketed Co2 Incubator LX400AJI is designed with an intelligent touchscreen controller and displays working parameters, date, time, time performance curve, humidity (optional), and CO2 concentration. It is available with a heat decontamination function with a HEPA high efficient filter. Equipped with temperature control system, door heating system, environment temperature detect system, over temperature protection system, and Power off the alarm system.

Specifications

Volume
155 L
Chamber dimension (W x D x H)
480 x 530 x 610 mm
Heating power
Air jacket microcomputer PID control
Temperature control range
RT+3 - 50℃
Work environment temp
+5 - 30℃
Temperature accuracy
±0.1℃
CO2 Control range
0 - 20%
CO2 Control accuracy
±0.1% (IR sensor)
CO2 Restore time
(Door open 30s, recovery to 5%) ≤ 3min
Temperature restore time
(Door open 30s, recovery to 37℃) ≤ 8min
Related humidity
Nature evaporation > 90%
Sterilization
180° Centigrade and HEPA high efficient filter
Input power
700 W
Voltage
AC220V/50Hz
Overall dimension (W x D x H)
670 x 767 x 880 mm
Packing dimension (W x D x H)
850 x 800 x 1070 mm
Gross weight
130 Kg
Features
  • Air jacketed heater
  • Standard sterilization method is 180° moist heat disinfection with HEPA high efficient filter
  • Built-in LCD touch screen display with controller
  • 72-hour machine operation record query function
  • Infrared sensors can keep CO2 concentration stable and uniform when doors open frequently
  • IR sensor is very sensitive to CO2 concentration varies and is measured accurately, if open the door is for 30s and close the door, is within 3 min the CO2 concentration can resume to the set value of 5%
  • Equipped with PT100 temperature sensor
  • Auto-controller of fan speed to prevent sample damage
  • Doors temperature controller prevents dewfall on the glass doors
  • Separate temperature-limiting alarm system for safe usage
  • It has an RS-485 communication port as options can be remotely controlled on a computer for monitoring the running and starting or close the machine
  • Designed with several safety features such as high and low temperature and over-temperature alarms, door open too long alarm, door temperature sensor failure alarm, chamber sensor failure alarm, CO2 condensation too high or too low alarm, disinfection and sterilization status reminder, independent temperature limiter alarm, power off alarm, over temp sensor failure alarm
Applications
  • Used in tissue culture, in vitro fertilization, Neuroscience, Cancer research, Stem cell research, Regenerative medicine, Mammalian cell research, etc.
Standard Accessories
  • Shelves: 3 Pcs
Optional Accessories
  • Only humidity display
  • Humidity control system (with humidity display)
  • RS485/232 Connector
  • CO2 Pressure-releasing valve